Autoclave 23 litres class B with USB
Manufacturer Information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Responsible person information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Warnings and product safety information:
Autoclave 23 litres class B with USB and liquids programme.
Programmes:
Programmes | Temperature | Pressure | Sterilising Time | Time total | Type | Max load | Max Load per tray |
---|---|---|---|---|---|---|---|
Solid | 134 ºC | 210 kPa | 4 min | 25-45 min | Solid unwrapped material | 5 Kg | 1.5 Kg |
121 ºC | 110 kPa | 20 min | 30-60 min | ||||
Wrapped | 134 ºC | 210 kPa | 4 min | 35-60 min | Solid unwrapped material | 5 Kg | 1.5 Kg |
121 ºC | 110 kPa | 20 min | 35-65 min | Solid or hollow material with simple wrap | 4 Kg | 1.2 Kg | |
Liquids | 134ºC | 210 kPa | 10 min | 30-55 min | Liquid | 1.2 Kg | 0.4 Kg |
121ºC | 110 kPa | 30 min | 35-60 min | ||||
Textile | 134 ºC | 210 kPa | 8 min | 45-65 min | Porous unwrapped material | 1.25 Kg | 0.4 Kg |
Porous material simple wrapped | 1.1 Kg | 0.3 Kg | |||||
Porous material double wrapped | 0.75 Kg | 0.25 Kg | |||||
121 ºC | 110 kPa | 30 min | 50-75 min | Hollow material simple wrapped | 4 Kg | 1.25 Kg | |
Solid or hollow material double wrapped | 2 Kg | 0.6 Kg | |||||
Prion | 134 ºC | 210 kPa | 18 min | 50-75 min | Porous unwrapped material | 1.25 Kg | 0.4 Kg |
Porous material simple wrapped | 1.1 Kg | 0.3 Kg | |||||
Porous material doble wrapped | 0.75 Kg | 0.25 Kg | |||||
Hollow material simple wrapped | 4 Kg | 1.25 Kg | |||||
Solid or hollow material doble wrapped | 2 Kg | 0.6 Kg | |||||
Test B&D | 134 ºC | 210 kPa | 3,5 min | 28-35 min | - | - | - |
Test Helix | 134 ºC | 210 kPa | 3,5 min | 28-35 min | - | - | - |
Vacuum Test | - | - | - | 20-25 min | - | - | - |
- The minimum time required for the autoclave to be ready for use after connecting to power is 5 minutes.
- The max. sterilisation cycle temperature of 134 ºC is 137 ºC
- The max. sterilisation cycle temperature of 121 ºC is 124 ºC
Features
- Fully automatic with preset sterilisation cycles.
- Chamber and door in stainless steel with electro-polished finish.
- High precision control system for perfect sterilisation results.
- User friendly interface.
- LCD screen displays information about the process continuously.
- Open type water tank, this feature makes it easy water filling and cleaning the tank.
- Languages: Spanish, English, German, French, Italian, Dutch, Polish, Romanian, Czech, Latvian, Lithuanian, Hungarian and Chinese.
Safety devices
- Breaker
Protects your equipment from any possibility of failure in the power supply.
Action: Interrupts the power supply. - Thermal circuit breaker
Protection against short circuits and possible overheating in the primary network of the transformer.
Action: Temporary interruption (until the temperature has dropped) in the winding. - Safety valve
Protection against a possible increase in pressure within the chamber.
Action: Escape steam until the restoration of a normal pressure level. - Micro switch to control de state of the door.
Exercises control in adjusting the door to start a sterilisation cycle.
Action: Transmits signal to alert the incorrect position of the door. - Thermostat for heating elements in the steam generator.
Protection againts possible overheating in the heating elements in the steam generator.
Action: Interrupts the power supply to the heating elements in the steam generator. - Thermostat for heating elements in the chamber.
Protection against possible overheating in the heating elements in the chamber.
Action: Interrupts the power supply to the heating elements in the chamber. - Auto leveling hydraulic system.
The hydraulic system returns to its natural if the sterilisation cycle is interrupted manually.
Action: The internal pressure is restored automatically. - Door lock
Effective to prevent accidental opening of the door protection.
Action: Prevents door opening by a hydraulic locking system.
Technical features:
- Nominal Voltage: 220-240V; 50/60 Hz; 10A
- Chamber dimensions: 247 x 450 mm
- Breaker: F16A / 400V
- Nominal power: 2200VA
- Sterilisation temperature: 121 ºC / 134 ºC
- Distilled water tank capacity: 2.5 L approx.
- Operating temperature: 5 ºC / 40 ºC
- Operating relative humidity: Max. 80%, non-condensing
- External dimensions: 480 x 455 x 700 mm
- Net weight: 51 Kg
- Max. noise level: <70 dB
- Operating atmospheric pressure: 76 / 106 kPa
Components:
- Distilled water tank
- LCD display
- Control panel
- Main switch
- Key for distilled and used water drain
- USB port
- Safety valve
- Ventilation grid
- Lock system
- Power cable
- Label
Accessories included:
- Steriliser
- 3 Instrument trays
- Base for trays
- Tray extraction handle
- Door adjust tool
- 2 drain hoses
- Door gasket
- Instruction manual
Optional accessories (not included):
- Ref. 922-PRINTER04: printer.
Autoclave - step by step:
1- Open the door, take the tools and accessories that are within the sterilization chamber, remove the wrapper and clean them.
2- Connect the power cord to a suitable power source.
3- Swith on: The main switch is located on the right bottom area in the front panel.
Once activated, the LCD screen is on, where basic operation icons such as the position of the door, the water level, the sterilisation programme, date, time, etc. appear.
4- Fill tank with distilled water.
Remove the top lid and fill, if you hear a signal, it means that you exceeded the permitted level of water in the tank, please stop immediately and drain the amount of water necessary for the signal off.
5- Preparation of material to be sterilised:
Place instruments that are of different materials in different trays, however take care to separate them before sterilisation.
If the instruments are not stainless steel, put some sterilisation paper between the tray and the instruments.
Do not allow direct contact between instruments of different materials.
Verify that all instruments are in correct position.
Place instruments such as lenses, containers, tubes, etc. inverted, so that water does not stagnate within them.
Do not overload the trays above their limit.
Do not overlap the trays on top of each other, or put them in direct contact with the wall of the sterilisation chamber.
Do not use metal clips or pins, this may impair the operation of the steriliser.
6- Select the sterilisation programme.
7- Start the sterilisation cycle.
After selecting the sterilisation programme, insert instruments into the chamber on the trays, using the supplied plastic handle.
Once the instruments are within the chamber, close the door.
Caution: Be sure to close the door, otherwise the steriliser displays an alarm and can not start or stops the programme during the cycle.
8. Start the sterilisation programme.
Press the start button, then the steriliser starts operating during approximately between 30-75 minutes.
9-After the sterilisation cycle.
After the pressure drops to zero, you can open the door and remove the instruments. The steriliser alerts by a beep once the sterilisation cycle is completed.