18 liter class B green autoclave
Description
Autoclave for sterilizing a wide variety of load types. 18-liter capacity, Class B, double safety lock, and USB port. Finished in a vibrant green color.
Programs:
- Solid
- Wrapped
- Liquid
- Textile
- Prion
- B&D Test
- Helix Test
- Vacuum test
Characteristics
- Fully automatic with pre-programmed sterilization cycles.
- Chamber and door in stainless steel, with electro-polished finish.
- High-precision control system for perfect sterilization results.
- User-friendly interface.
- LCD screen, continuously displays information about the cycle process.
- Open-type water tank. This feature makes filling and cleaning the tank easier. IMPORTANT: Do not reuse the water from the waste tank.
- Optional printer (not included. Reference 922-PRINTER04 ).
- Languages: Spanish, English, German, French, Italian, Portuguese, Dutch, Polish, Romanian, Czech, Latvian, Lithuanian, Hungarian and Chinese.
Security devices
- Circuit breaker.
Protect the equipment from any possibility of power supply failure.
Action: Interrupt the power supply. - Thermal circuit breakers.
Protection against possible short circuits and possible overheating in the primary network of the transformer.
Action: Temporary interruption (until the temperature has dropped) in the winding. - Safety valve.
Protection against a possible increase in pressure inside the chamber.
Action: Steam release until a normal pressure level is restored. - Micro switch to control the door status.
It controls the door adjustment to start a sterilization cycle.
Action: Transmits the signal to alert about the incorrect position of the door. - Thermostat for the heating elements in the steam generator.
Protection against possible overheating of the heating elements in the steam generator.
Action: Interrupts the power supply to the heating elements in the steam generator. - Thermostat for the heating elements in the chamber.
Protection against possible overheating of the heating elements in the chamber.
Action: Interrupts the power supply to the heating elements in the chamber. - Automatic leveling of the hydraulic system.
The hydraulic system will return to its natural pressure if the sterilization cycle is manually interrupted.
Action: Internal pressure will be restored automatically. - Door lock
Effective protection to prevent accidental opening of the door.
Action: Prevents the door from opening using a hydraulic locking system.
Technical specifications:
- Nominal Voltage: 220-240V; 50/60 Hz; 10A
- Camera dimensions: 247 x 350 mm
- Circuit breaker: F16A / 400V
- Nominal Power: 2200VA
- Sterilization temperature: 121 ºC / 134 ºC
- Distilled water tank capacity: approx. 2.5 L
- Operating temperature: 5 ºC / 40 ºC
- Operating relative humidity: Max. 80%, non-condensing
- External dimensions: 480 x 455 x 600 mm
- Net Weight: 47 Kg
- Maximum noise level: <70 dB
- Atmospheric operating pressure: 76 / 106 kPa
Components:
- Distilled water tank.
- LCD screen.
- Control Panel.
- Main switch.
- Tap for draining distilled and used water.
- USB port.
- Safety valve.
- Ventilation grille.
- Locking system.
- Power cable.
Included accessories:
- Instrument trays (3).
- Base for trays.
- USB flash drive.
- Handle for removing the trays.
- Door adjustment tool.
- Drainage hoses (2).
- Door seal.
Optional accessories and spare parts:
- 922-PRINTER04 Autoclave printer.
Autoclave process step by step:
1- Open the door, take the instruments and accessories inside the sterilization chamber, remove the wrapping and proceed to clean them.
2- Connect the power cord to a suitable power source.
3- Connect the printer ( optional accessory, sold separately. If you wish to purchase it, look for reference 922-PRINTER04 on this website ).
4- Power On: The main switch is located on the lower right front of the equipment.
- Once activated, the LCD screen will light up, displaying the basic operating icons of the sterilizer, such as door position, water level, sterilization program, date, time, etc.
5- Fill the tank with distilled water.
- Remove the top cover of the tank and proceed to fill it. If you hear a signal, it means you have exceeded the permitted water level in the tank. Please stop immediately and drain the necessary amount of water to turn off the signal.
6- Preparation of the material to be sterilized:
- Place instruments made of different materials in separate trays, but be sure to separate them before sterilization.
- If the instruments are not made of stainless steel, place sterilizing paper between the tray and the instruments.
- Do not allow direct contact between instruments made of different materials.
- Verify that all instruments are in the correct position.
- Place instruments such as lenses, containers, tubes, etc. in the tray upside down, so that water does not stagnate inside them.
- Do not overload the trays beyond their limit.
- Do not stack the trays on top of each other, nor place them in direct contact with the wall of the sterilization chamber.
- Do not use metal clips or pins, this may jeopardize the functioning of the sterilizer.
7- Select the sterilization program.
8- Start the sterilization cycle.
- After selecting the sterilization program, place the instruments into the chamber using the trays and the supplied plastic handle.
- Once the instruments are inside the chamber, proceed to close the door.
- Caution: Make sure to close the door properly, otherwise the sterilizer will display an alarm and will not be able to start the program or will stop during the cycle.
9- Start the sterilization program.
- Press the start button, then the sterilizer will start working for approximately 30-75 minutes.
10-Sterilization cycle completed.
- Once the sterilization cycle is complete, the printer ( optional accessory with reference 922-PRINTER04 ) will automatically print the complete cycle report.
- After the pressure drops to zero, you can open the door and remove the instruments. The sterilizer will alert you with an audible signal once the sterilization cycle is complete.
- Caution: Whenever removing instruments, use the plastic handle supplied for the tray holder.
IMPORTANT: Do not reuse water from the waste tank.
Facility:
- There must be at least 10 cm of free space around the sterilizer, 25 cm at the top, and enough space for the door to open completely.
- The area where the sterilizer operates must have adequate ventilation.
- The sterilizer should be placed on a weight-resistant table at a suitable level for handling.
- Do not cover or block the side vents or the condenser fan air outlet grille.
- Do not place the sterilizer near water sources where it may be splashed.
- Keep it away from other heat sources.
Example of a printed report * :
* The printer is an optional accessory and is sold separately with reference 922-PRINTER04 .
================================
Program: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying time: 08Min
Sterilization time: 4.0 min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -080.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -080.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature: 135.2C
Min. Temperature: 134.3C
Max. Pressure: 214.0kPa
Min. Pressure: 204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
Cycle No.: 00017
Result: success
Date: 2014-06-07
SN:E54723
Operator:
================================
Regulations:
- 93/42/EEC Medical Health Directive.
- 97/23/EEC Pressure Equipment Directive.
- EN 13060 Small water steam sterilizers.
- EN 61326, EN 61000 Electromagnetic compatibility.
- EN 61010-1, UL 61010-1 Safety requirements.
- EN 61010-2-040, IEC 61010-2-040 Specific requirements for steam sterilizers used for the treatment of medical materials.
Video: