Vital signs monitor ECG, SpO2, NIBP, RESP and TEMP
Manufacturer Information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Responsible person information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Warnings and product safety information:
This vital signs monitor is intended to be used in laboratories and hospital or clinic areas, where monitoring systems are needed. It provides a fast and reliable measurement for both adult and newborn patients.
Features:
Heart rate / pulse: is measured with the ECG, oximetry, and noninvasive blood pressure (NIBP) parameters.
Electrocardiogram (ECG):
- The monitor provides continuous 5-lead ECG processing, with standard derivation selections and electrocautery discharge filtering.
- You can display up to 3 waveform channels simultaneously.
Oximetry: The oximetry parameter provides continuous noninvasive monitoring of oxygen saturation (% SpO2) in the arterial and pulse (PR) rates.
Non-invasive blood pressure (NIBP):
- The noninvasive blood pressure (NIBP) parameter provides the values of systolic, diastolic, and mean arterial pressure, in addition to the heart rate.
- NIBP measurements can be made in automatic, manual or STAT modes.
Respiratory frequency (RR): The monitor provides the respiratory frequency and the Resp waveform by measuring the impedance between 2 ECG electrodes at the patient's thorax.
Temperature (TEMP): There are two independent temperature control channels (T1 and T2).
Display:
- The monitor has a high resolution color LCD screen and high contrast.
- Provides a continuous, real-time image of up to 4 waveforms.
- It also shows the measured values, chronological data, measurement trends, alarm limits and patient information.
Internal battery:
- The rechargeable internal Li battery is Mainly intended for backup and usage switching.
- The expected lifespan of the battery is 2 to 5 years, depending on usage.
- A fully charged battery will last 2 to 3.5 hours, depending on the use of the monitor.
Technical specifications:
Display:
- 7 inch high resolution, active matrix flat screen (TFT or Thin film transistor).
- Resolution: 800 × 480 pixels.
Indicators:
- Work status LED
- AC power LED
- Battery charge LED
- Battery power LED
- Alarm silence LED
Alarm volume: from 45dBA to 85 dBA, 1 meter away (adjustable).
ECG:
- Heart rate range: 20-350 bpm
- Heart Rate Accuracy: 2 bpm or ± 2 % (whichever is greater)
- QRS detection range: 0.5 to 5 mV
- Pacemaker pulse detection: Amplitude: 2 mV to 700 mV
- Detection of duration: 0.1 - 2.0
- Heart rate range alarm:
- High: 20-350 bpm and deactivated
- Low: 20-350 bpm and deactivated
- Average of heart rate: Average of 8 seconds set
- Selection of leads: I, II, III, V, aVR, aVL, or aVF (5 leads)
- Gain display settings: X1/4, X1/2, X1, X2, X4
- Input range: -5,0 mV to +5,0 mV
- Frequency response: 0.05 Hz to 40 Hz
- Input impedance: Differential > 5 Mohms, as required and tested by ANSI / AAMI EC-13.
- Leaks: < 10 uA
- Patient isolation: > 4000 VCA
- Visualisation of waveforms: 120 Hz; 6.25, 12.5, 25, or 50 mm/s
- Update of frequency digits: 1 Hz
SpO2:
- Range of SpO2: 0-100 % of functional saturation
- SpO2 precision:
- ± 2% @ 70-100 %
- <70% unspecified
- Frequency range of the pulse: 20-350 bpm
- Frequency precision of the pulse: ± 2 bpm or ± 2 % (whichever is greater)
- Alarm ranges of SpO2:
- High: 50 - -100 % and deactivated
- Low: 50 - -100 % and deactivated
- Calbration: SpO2 default calibration on 50 % to 100 % range, using human blood samples for the functional saturation.
NIBP:
Measurement of blood pressure:
- Method of measurement: Oscillometric with reducing deflation
- Range: Systolic: 40 to 270 mmHg
- Average arterial: 30 to 220 mmHg
- Diastolic: 20 to 200 mmHg
- Pulse: 25 to 300 bpm
- NIBP precision: Algorithm based on the human algorithm that meets or exceeds the precision requirements of the norms ANSI/AAMI SP10:1992 and 2002 for non-invasive arterial pressure through the oscillometric method.
Configuration of inflation pressure:
- Time of measuring: 30 to 50 seconds (typical), 120 seconds (maximum)
- Predefined inflation pressure:
- 120 mmHg - Adult
- 100 mmHg - Paediatric
- 90 mmHg - Neonatal
- Calibration: Calibration at the factory
- AUTO interval times: 2, 3, 5, 10, or 30 minutes, or 1, 2 hours 20-260 (in 1 mmHg steps), and deactivated
- Alarm range: 25- 250 lpm, ± 2% or ± 3 lbpm (whichever is greater)
- Pulse frequency precision: 251-300 bpm, ± 5 %
Respiratory frequency (RESP):
- Range: 0-120 breaths per minute (bpm)
- Precision: ±1 bpm
- Resolution: 1 bpm
- RESP alarm range:
- High: 0-120 bpm (1 bpm steps) and deactivated
- Low: 0-120 bpm (1 bpm steps) and deactivated
Temperature (TEMP):
- Channel: Two
- Range: 25-45ºC
- Precision:± 0.2ºC plus the tolerance of the temperature sensor
- Resolution: 0.1ºC
- Alarm range of TEMP:
- High: 35.5-43.5ºC and deactivated in increments of 0.1
- Low: 35.5-43.5ºC and deactivated in increments of 0.1
Voltage requirements: AC entry: 100 to 240 V, 50/60 Hz
Dimensions:
- Width: 300 mm.
- Length: 180 mm.
- Height: 130 mm.
- Equipment weight: 2 Kg approx.
- Total weight: 3.6 Kg approx.
Environmental protection:
- Temperature:
- From 0 to 50 0C (operation)
- From -40 to 75 0C (storaging)
- Humidity:
- From 15 to 95 % (operation)
- From 10 to 95 % (storaging)
Classification of the equipment:
- Type of protection (from electrocutions): TYPE I (internal power)
- Operational mode: Continuous
- Protection degree (from liquids absorption): IPX1, drop-proof
- Mobility degree: Portable
- Protection degree (from electrocutions): CF type
- Safety requirements: EN60601-1:2002
Includes:
- Vital signs monitor
- ECG cable 5 leads
- Temperature probe
- Reusable finger oximetry sensor for adults
- Tube for NIBP with sleeve for adults 27-35 cm
- Internal battery
- Power cord
- Intructions manual
Certification:
- CE 93/42 directive of Health Products