Autoclave 18 litres class B with USB
Manufacturer Information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Responsible person information:
Name or company: Pending update
Email: Pending update
Postal address: Pending update
City: Pending update
Country: Pending update
Warnings and product safety information:
Autoclave 18 litres Class B with USB.
Reference | Printer | Programme liquids |
922-STE-B18D | No (optional accessory, it is sold separately. If you wish to purchase it, look for the reference 922-PRINTER04). | Yes |
Programmes:
Programmes | Temperature | Pressure | Sterilising time | Overall time | Type | Max load | Max load per tray |
Solid | 134 ºC | 210 kPa | 4 min | 20-35 min | Solid unwrapped material | 4.5 Kg | 1.2 Kg |
121 ºC | 110 kPa | 20 min | 30-50 min | ||||
Wrapped | 134 ºC | 210 kPa | 4 min | 30-50 min | Solid unwrapped material | 4.5 Kg | 1.2 Kg |
121 ºC | 110 kPa | 20 min | 35-55 min | Solid or hollow material simple wrapped | 3.5 Kg | 1.1 Kg | |
Liquids | 134ºC | 210 kPa | 10 min | 30-55 min | Liquid | 1 Kg | 0.3 Kg |
121ºC | 110kPa | 30 min | 35-60 min | ||||
Textile | 134 ºC | 210 kPa | 8 min | 40-65 min | Porous unwrapped material | 1 Kg | 0.3 Kg |
Porous material simple wrapped | 0.80 Kg | 0.25 Kg | |||||
Porous material double wrapped | 0.6 Kg | 0.2 Kg | |||||
121 ºC | 110 kPa | 30 min | 45-75 min | Hollow material simple wrapped | 3.5 Kg | 1.25 Kg | |
Solid or hollow material double wrapped | 1.5 Kg | 0.6 Kg | |||||
Prion | 134 ºC | 210 kPa | 18 min | 45-70 min | Porous unwrapped material | 1 Kg | 0.3 Kg |
Porous material simple wrapped | 0.75 Kg | 0.25 Kg | |||||
Porous material double wrapped | 0.6 Kg | 0.2 Kg | |||||
Hollow material simple wrapped | 3.5 Kg | 1 Kg | |||||
Solid or hollow material double wrapped | 1.5 Kg | 0.5 Kg | |||||
Test B&D | 134 ºC | 210 kPa | 3.5 min | 22-35 min | - | - | - |
Test Helix | 134 ºC | 210 kPa | 3.5 min | 22-35 min | - | - | - |
Vacuum test | - | - | - | 15-20 min | - | - | - |
- The minimum time required autoclave to be ready for use after connecting is 5 minutes.
- The max. temperature for the sterilisation cycle of 134 ° C is 137 ºC
- The max. temperature for the sterilisation cycle of 121 ° C is 124 ° C
Features:
- Fully automatic with pre-programmed sterilisation cycles.
- Chamber and door in stainless steel with electro-polished finish.
- High precision control system for perfect sterilisation results.
- User friendly interface.
- LCD screen displays information about the process continuously.
- Open type water tank, this feature makes it easy water filling and cleaning the tank.
- Optional printer (not included. Reference 922-PRINTER04).
- Languages: Spanish, English, German, French, Italian, Dutch, Polish, Romanian, Czech, Latvian, Lithuanian, Hungarian and Chinese.
Safety devices:
- Breaker.
- Protect your computer from any possibility of failure in the power supply.
- Action: Interrupt the power supply.
- Thermal circuit breaker.
- Protection against short circuits and possible overheating in the transformer primary network.
- Action: Temporary interruption (until the temperature has dropped) in the winding.
- Safety valve.
- Protection against a possible increase in pressure within the chamber.
- Action: Escape steam until the restoration of a normal pressure level.
- Micro switch to control the state of the door.
- Exercises control in adjusting the door to start a sterilisation cycle.
- Action: Transmits signal to alert the incorrect position of the door.
- Thermostat for heating elements in the steam generator.
- Protection against possible overheating in the heating elements in the steam generator.
- Action: interrupts the power supply to the heating elements in the steam generator.
- Thermostat for heating elements in the chamber.
- Protection against possible overheating in the heating elements in the chamber.
- Action: interrupts the power supply to the heating elements in the chamber.
- Auto leveling hydraulic system.
- The hydraulic system pressure returns to its natural if the sterilization cycle is interrupted manually.
- Action: the internal pressure is restored automatically.
- Door Lock
- Effective to prevent accidental opening of the door protection.
- Action: Prevents door opening by a hydraulic locking system.
Technical features:
- Nominal Voltage: 220-240V; 50/60 Hz; 10A
- Chamber dimensions: 247 x 350 mm
- Breaker: F16A / 400V
- Nominal power: 2200VA
- Sterilisation temperature: 121 ° C / 134 ° C
- Capacity distilled water tank: 2.5 L approx.
- Operating temperature: 5 ° C / 40 ° C
- Operating relative humidity: Max. 80%, non-condensing
- External dimensions: 480 x 455 x 600 mm
- Net weight: 47 Kg
- Max noise level: <70 dB
- Operating atmospheric pressure: 76/106 kPa
Components:
- Distilled water tank
- LCD display
- Control panel
- Main switch
- Drain valve for distilled and used water
- USB port
- Safety valve
- Ventilation grid
- Lock system
- Power cable
- Rating plate
Accessories included:
- Steriliser
- Three trays of instruments
- Base for trays
- Extraction handle
- Tool for door adjustment
- 2 drain hoses
- Door gasket
- User manual
Optional accessories and spare parts:
- Printer for autoclave (ref. 922-PRINTER04) and paper roll for autoclave printer (ref. 534-ROLLPAPER).
Autoclave process step by step:
1- Open the door, take the tools and accessories that are within the sterilisation chamber, remove the wrapper and clean them.
2- Plug the power cord to a suitable power source.
3- Plug the printer (optional accessory, it is sold separately. If you wish to purchase it, please look the reference 922-PRINTER04).
4- Switch on: The main switch is located on the right bottom in the front panel.
Once activated, the LCD screen is on, where basic operation icons of the steriliser appear, such as the position of the door, the water level, the sterilisation programme, date, time, etc.
5- Fill tank with distilled water.
Remove the top lid and fill, if you hear a signal, it means that you exceeded the permitted level of water in the tank, please stop immediately and drain the amount of water necessary for the signal off.
6- Preparation of material to be sterilised:
Place instruments that are of different materials in different trays, however take care to separate them before sterilisation.
If the instruments are not stainless steel, put some sterilisation paper between the tray and the instruments.
Do not allow direct contact between instruments of different materials.
Verify that all instruments are in correct position.
Place on tray instruments such as lenses, containers, tubes, etc. inverted, so that water does not stagnate within them.
Do not overload the trays above their limit.
Do not overlap the trays on top of each other, or put them in direct contact with the wall of the sterilisation chamber.
Do not use metal clips or pins, this may impair the operation of the steriliser.
7- Select the sterilisation programme.
8- Start the sterilisation cycle.
After selecting the sterilisation programme, insert instruments into the chamber through the trays, using the supplied plastic handle.
Once the instruments are within the chamber, close the door.
Caution: Be sure to close the door, otherwise the steriliser displays an alarm and can not start or stop the programme during the cycle.
9. Start the sterilisation programme.
Press the start button, then the steriliser start operating during approximately between 30-75 minutes.
10-After the sterilisation cycle.
After completion of the sterilisation cycle, the printer will print the full report automatically (optional accessory with reference 922-PRINTER04)
After the pressure drops to zero, you can open the door and remove the instruments. The steriliser alerts by a beep once the sterilisation cycle is completed.
Caution: Whenever you remove instruments, use the supplied plastic handle for supporting trays.
IMPORTANT: Do not reuse waste water.
Installation:
- There must be at least 10 cm of free space around the sterilizer, 25 cm at the top, and enough room for the door to open.
- The place where you use the steriliser must have adequate ventilation.
- The steriliser must be placed on a stable surface and at an adequate height for handling.
- Do not cover or block the side outlets or outlet grille air condenser fan.
- Do not place the steriliser near water sources where it can be splashed.
- Keep away from heat sources.
Example of a printed report:
* The printer is an optional accessory and is sold separately. If you wish to purchase it, please look for the reference 922-PRINTER04.
================================
Programme: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying time: 08min
Sterilisation time: 4.0Min
--------------------------------
Time Temp. pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -080.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -080.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature: 135.2C
Min. Temperature: 134.3C
Max. Pressure: 214.0kPa
Min. Pressure: 204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
13:04:07 End 102.4C
--------------------------------
Cycle No .: 00017
Result: success
Date: 07/06/2014
SN: E54723
operator:
================================
Regulations:
- 93/42/EEC Medical Device Directive.
- 97/23/EEC Pressure Equipment Directive.
- EN 13060 Small steam water steriliser.
- EN 61326, EN 61000 Electromagnetic compatibility.
- EN 61010-1, UL 61010-1 Safety requirements.
- EN 61010-2-040, IEC 61010-2-040 Specific requirements for steam sterilisers used for treatment of medical materials.
In the UK this product needs to be inspected and certified anually. Quirumed don´t offer this service and you will need to find your own local techncian who can do this. We do however have a technical service based in Spain to cover the product during the warranty period
Video: